1. Meeting Overview ​

ObjectiveThis review checks the completeness and consistency of all documentation and product-related activities before release of the in-house software.
LocationMS Teams
DateDay-Month-Year
AddendumDay-Month-Year

The meeting on Day-Month-Year took place while the product was in a pre-final state. Completion of the outstanding items was communicated to all participants by email on Day-Month-Year as an addendum.

Items addressed:

  1. Inclusion of product name in the local IT maintenance windows (appointment agreed)
    Processing: appointment on Day-Month-Year and inclusion completed.

  2. Automated test documentation for approximately 70 programmed tests outstanding (appointment already planned)
    Processing: completed; see test documentation.

  3. One risk control measure (schemas for the ETL process) was implemented only partially. Due to the very low risk resulting from the missing measure, the decision was made to perform a risk-benefit assessment and implement the measure in version 1.1. (appointment already planned)
    Processing: risk-benefit assessment appointment completed, additional tests for the ETL path created, schemas planned for version 1.1.

2. Product ​

Product nameVersion
Product name1.0

3. Participants ​

NameAffiliationRole / Responsibility
Project lead namelocal development department, Hospital NameProject lead product name
Data processing expert namelocal development department, Hospital Nameproduct name staff, especially technical documentation and QM
Medical and risk expert namelocal development department, Hospital Nameproduct name staff, especially risk management and QM
Technical tester namelocal development department, Hospital Nameproduct name staff
Quality management reviewer namelocal quality management departmentReview of the risk management report
Risk management reviewer namelocal risk management departmentReview of the risk management report
Medical technology reviewer namelocal medical technology departmentReview of the risk management report
Information management reviewer namelocal information management departmentReview of the risk management report
External consultant nameexternal consulting organizationConsulting
Legal consultant namelocal legal departmentConsulting

4. Documentation Checklist ​

The following documents are up to date, complete, and approved:

Document nameYes/NoComment / open activities
Product specificationYesintended-use
Product classificationYesmdr-classification
Development planYesdevelopment-plan
Additional development schedule: development-chart
Stakeholder requirementsYesstakeholder-requirements
Stakeholder requirements reviewYesstakeholder-requirements-review
System requirementsYessystem-requirements
System requirements reviewYessystem-requirements-review
Architecture and designYesarchitecture
SOUP listYessoup-list
Architecture and design reviewYesarchitecture-review
Test planYestest-plan
Test specificationYestest-specifications
Test protocol(s)Yestest-protocol-UIT
test-protocol-ST
test-protocol-AT
Test report(s)Yestest-report
Deployment procedureYesdeployment
Instructions for useYesinstructions-for-use
Bug-fix listNoFirst product version, not applicable
List of known anomaliesNoFirst product version, not applicable
Risk management planYesrisk-management-plan
Risk analysis tableYesrisk-table
Risk management reportYesrisk-management-report
Market analysisYesmarket-analysis
Public declaration for in-house softwareNoNot applicable. The central hospital document is not part of this local publication.
Medical device inventory for in-house softwareNoNot applicable. Replaced by activity "Entry in the local medical device inventory" in section 5
Release notesNoFirst product version, not applicable

5. Activities Performed ​

The following activities are complete:

ActivityYes/NoComment
Specification of all relevant product functionalities; completion and review of requirementsYesSee stakeholder requirements plus traceability in system requirements and architecture and design
Evaluation of all relevant risks, including risks from known anomaliesYesSee risk analysis table, especially product FMEA.

Addendum: review of the risk management report completed
Verification and validation activities, i.e. all test phasesYesTests fully performed.

Addendum: acceptance test with pharmacists completed on Day-Month-Year, documentation completed in test protocols and test report
Complete documentation of traceability: stakeholder requirements can be assigned to system requirements and acceptance tests; system requirements can be assigned to architecture components and system tests; architecture components can be assigned to implemented software units and unit/integration tests; all risks and risk control measures can be assigned to resulting requirementsYes1. Stakeholder -> system: Yes; stakeholder -> acceptance test: pre-final.

Addendum: acceptance test completed on Day-Month-Year.

2. System -> architecture: Yes; system -> system tests: pre-final.

Addendum: test documentation completed.

3. Architecture -> software units: Yes; architecture -> unit/integration tests: pre-final.

Addendum: test documentation completed.

4. Risk measures -> requirements: Yes
Assignment of and reference to the version number, as defined in the development plan, in all product-related documentsYes
Entry in the local medical device inventoryYesRegistration form submitted and local medical device ID assigned on Day-Month-Year

6. Result ​

Product release is granted.

As of Day-Month-Year, not all release conditions had been fulfilled. The participants agreed that transmission of the outstanding content would be a condition for release.

Outstanding content was transmitted on Day-Month-Year. Review of the final outstanding document, the risk management report, was completed on Day-Month-Year.

7. Applicable Documents ​

Document nameReference (local ID)