Release Checklist for In-House Software - LAMPE
| Template ID | Local ID |
| Template Version | Version |
| Approval | Approver name |
1. Meeting Overview
| Objective | This review checks the completeness and consistency of all documentation and product-related activities before release of the in-house software. |
| Location | MS Teams |
| Date | Day-Month-Year |
| Addendum | Day-Month-Year |
The meeting on Day-Month-Year took place while the product was in a pre-final state. Completion of the outstanding items was communicated to all participants by email on Day-Month-Year as an addendum.
Items addressed:
Inclusion of product name in the local IT maintenance windows (appointment agreed)
Processing: appointment on Day-Month-Year and inclusion completed.Automated test documentation for approximately 70 programmed tests outstanding (appointment already planned)
Processing: completed; see test documentation.One risk control measure (schemas for the ETL process) was implemented only partially. Due to the very low risk resulting from the missing measure, the decision was made to perform a risk-benefit assessment and implement the measure in version 1.1. (appointment already planned)
Processing: risk-benefit assessment appointment completed, additional tests for the ETL path created, schemas planned for version 1.1.
2. Product
| Product name | Version |
|---|---|
| Product name | 1.0 |
3. Participants
| Name | Affiliation | Role / Responsibility |
|---|---|---|
| Project lead name | local development department, Hospital Name | Project lead product name |
| Data processing expert name | local development department, Hospital Name | product name staff, especially technical documentation and QM |
| Medical and risk expert name | local development department, Hospital Name | product name staff, especially risk management and QM |
| Technical tester name | local development department, Hospital Name | product name staff |
| Quality management reviewer name | local quality management department | Review of the risk management report |
| Risk management reviewer name | local risk management department | Review of the risk management report |
| Medical technology reviewer name | local medical technology department | Review of the risk management report |
| Information management reviewer name | local information management department | Review of the risk management report |
| External consultant name | external consulting organization | Consulting |
| Legal consultant name | local legal department | Consulting |
4. Documentation Checklist
The following documents are up to date, complete, and approved:
| Document name | Yes/No | Comment / open activities |
|---|---|---|
| Product specification | Yes | intended-use |
| Product classification | Yes | mdr-classification |
| Development plan | Yes | development-plan Additional development schedule: development-chart |
| Stakeholder requirements | Yes | stakeholder-requirements |
| Stakeholder requirements review | Yes | stakeholder-requirements-review |
| System requirements | Yes | system-requirements |
| System requirements review | Yes | system-requirements-review |
| Architecture and design | Yes | architecture |
| SOUP list | Yes | soup-list |
| Architecture and design review | Yes | architecture-review |
| Test plan | Yes | test-plan |
| Test specification | Yes | test-specifications |
| Test protocol(s) | Yes | test-protocol-UIT test-protocol-ST test-protocol-AT |
| Test report(s) | Yes | test-report |
| Deployment procedure | Yes | deployment |
| Instructions for use | Yes | instructions-for-use |
| Bug-fix list | No | First product version, not applicable |
| List of known anomalies | No | First product version, not applicable |
| Risk management plan | Yes | risk-management-plan |
| Risk analysis table | Yes | risk-table |
| Risk management report | Yes | risk-management-report |
| Market analysis | Yes | market-analysis |
| Public declaration for in-house software | No | Not applicable. The central hospital document is not part of this local publication. |
| Medical device inventory for in-house software | No | Not applicable. Replaced by activity "Entry in the local medical device inventory" in section 5 |
| Release notes | No | First product version, not applicable |
5. Activities Performed
The following activities are complete:
| Activity | Yes/No | Comment |
|---|---|---|
| Specification of all relevant product functionalities; completion and review of requirements | Yes | See stakeholder requirements plus traceability in system requirements and architecture and design |
| Evaluation of all relevant risks, including risks from known anomalies | Yes | See risk analysis table, especially product FMEA. Addendum: review of the risk management report completed |
| Verification and validation activities, i.e. all test phases | Yes | Tests fully performed. Addendum: acceptance test with pharmacists completed on Day-Month-Year, documentation completed in test protocols and test report |
| Complete documentation of traceability: stakeholder requirements can be assigned to system requirements and acceptance tests; system requirements can be assigned to architecture components and system tests; architecture components can be assigned to implemented software units and unit/integration tests; all risks and risk control measures can be assigned to resulting requirements | Yes | 1. Stakeholder -> system: Yes; stakeholder -> acceptance test: pre-final. Addendum: acceptance test completed on Day-Month-Year. 2. System -> architecture: Yes; system -> system tests: pre-final. Addendum: test documentation completed. 3. Architecture -> software units: Yes; architecture -> unit/integration tests: pre-final. Addendum: test documentation completed. 4. Risk measures -> requirements: Yes |
| Assignment of and reference to the version number, as defined in the development plan, in all product-related documents | Yes | |
| Entry in the local medical device inventory | Yes | Registration form submitted and local medical device ID assigned on Day-Month-Year |
6. Result
Product release is granted.
As of Day-Month-Year, not all release conditions had been fulfilled. The participants agreed that transmission of the outstanding content would be a condition for release.
Outstanding content was transmitted on Day-Month-Year. Review of the final outstanding document, the risk management report, was completed on Day-Month-Year.
7. Applicable Documents
| Document name | Reference (local ID) |
|---|---|
| See references in section 5. |