Studies ​

Why do studies on patient outcomes require a certified medical device? ​

A CDSS must be used in patient care to investigate patient outcomes. Consequently, the CDSS automatically meets the definition of a medical device within the EU and falls under the scope of the MDR.

This well-known requirement has led to a far less recognized consequence: there is remarkably little evidence on how medical device software impacts patient outcomes. That is because every "workaround study design" shares the same limitation: they cannot evaluate the impact of a CDSS in patient care, leaving them restricted to measuring theoretical proxies as endpoints (e.g. hypothetical case studies with healthcare professionals) or retrospective data analysis.

What is a clinical evaluation and what is a clinical investigation under the MDR? ​

Both are established terms under the MDR. A "clinical evaluation" involves generating, analyzing and assessing clinical data pertaining to a device or consulting scientific literature. It is always required before putting the product into service, even if it is not reviewed by a notified body for Class I devices or in-house medical devices. A "clinical investigation" refers to prospective studies that gather product-specific data. Requirements vary depending on the risk class:

Risk ClassClinical Investigation Required?
Class INo
Class IIa/IIbSometimes
Class IIIAlways


For a CDSS, the clinical evaluation without a clinical investigation is usually sufficient. Furthermore, requirements for the clinical evaluation of in-house medical devices are less extensive and formal. The guidance document MDCG 2023-1 outlines the following requirements for demonstrating the clinical performance of in-house devices:

"Performance data: according to Annex I of the IVDR/MDR, devices shall be designed and manufactured in such a way that they are suitable with regard to the performance they are intended to achieve, taking account of the generally acknowledged state of the art. A description of, where applicable, the analytical and the clinical performance data supporting the intended purpose should be provided."

Validating a CDS model with retrospective data and reporting statistical metrics (e.g., AUROC, sensitivity, specificity, PPV, NPV) is equivalent to the clinical evaluation standard for comparable CE-marked products.

What needs to be considered for interventional studies with in-house medical devices under the MDR? ​

It must be checked if the study is considered a "clinical investigation" under the MDR. If sufficient "analytical and clinical performance data supporting the intended purpose" is missing, the medical device is still "in development." In this case, any interventional study conducted with the device qualifies as a "clinical investigation." However, if sufficient data is already available (meaning the in-house medical device has reached the stage where it can be used in routine patient care) the classification depends on whether the device is being used within or outside its intended purpose (MDCG 2025-5, see also this article for details):

In-house medical devicesClinical investigation under MDR
Interventional study in developmentyes
Interventional study outside intended purposeyes
Interventional study within intended purposeno


For any interventional study of patient outcomes using an in-house CDSS, which is finalized and used within its intended purpose, the regulatory framework for a "clinical investigation" under the MDR no longer applies. If sufficient (retrospective) data is already available for the clinical evaluation, any further prospective evaluation conducted within its intended purpose falls outside the scope of a "clinical investigation". This significantly lowers regulatory hurdles, enabling a fast-track process for interventional studies with in-house medical devices.

What types of study designs exist, and how do their requirements vary? ​

In summary, possible study designs with clinical decision support systems can be divided into four groups:

No change in patient careChange in patient care
Historic data useRetrospective observational
(e.g. model development)
Before-and-after
(e.g. PMCF investigation)
New data collectionProspective observational
(e.g. silent testing)
Prospective interventional
(e.g. RCT / PCT)

PMCF: Post-Market Clinical Follow-up; RCT: Randomized Controlled Trial; PCT: Pragmatic Clinical Trial

The following simplified table provides an overview of the German requirements for each study design. Similar regulations apply across Europe. A German example for an national competent authority is the BfArM. For all study designs, early consultation with the ethics committee is recommended. Ultimately, the committee determines whether ethical approval is required (or a certificate of "no objection" is issued), as well as if and how informed patient consent must be obtained.

Retrospective observationalProspective observationalBefore-and-afterProspective interventional (in-house MD)"Clinical investigation" under MDR
Level of evidencelowmediumhighgold standardvariable
Medical device software requirednonoyesyesyes
Patient consent requiredunlikely¹unlikely¹unlikely¹likely¹likely¹
Patient insurance requirednononono²yes²
Permission of national competent authority requirednonononoyes³

¹If broad consent is available this often covers data use of retrospective studies. Exceptions for prospective interventional studies also exist, but require proof that consent procedures impose undue burden or compromise the scientific validity.
²A patient insurance is not required for "other clinical investigations" (Art. 82, MDR), provided that the medical device is used within its intended purpose and without any additional invasive or burdensome procedures. If the same conditions exist for a study with an in-house medical device, it can accordingly be assumed that the same exception (MPDG §26) applies.
³A simular exception (MPDG §47) also applies here.

Is a patient insurance required for interventional randomized studies with a in-house medical device within its intended purpose? ​

Within the domain of clinical decision support, interventions generally involve minimal risk. Ethics committees are often unfamiliar with this constellation in medical studies and might be assessing them too strictly. Here, the MDR serves as an effective benchmark. Given its rigorous approach to risk assessment, an MDR determination that a device does not pose sufficient risk to warrant measures such as mandatory patient insurance provides ethics committees with a widely recognized regulatory foundation for their evaluation.

Let's take a closer look at the MDR: It must be assumed that the requirements for a interventional study with an in-house medical device within its intended purpose and without any additional invasive or burdensome procedures will not exceed those set for "other clinical investigation" with a medical device that is used within its intended purpose and without any additional invasive or burdensome procedures. Following the assessment for comparable studies in the MDR a patient insurance should not be required here.

But wait, what is considered an"additional burdensome or invasive procedure"? Fortunately, this has been clarified. The most important point is that these procedures are "additional to those performed under the normal conditions of use of the device" (MDCG 2021-6). Therefore, if an invasive measure is already part of the device's standard usage (e.g., a diabetes alert triggering an insulin injection), and you simply randomize which patients receive the medical device (e.g. alerts) and which do not, this leads to no "additional burdensome or invasive procedures".

However, national regulations can differ. Early alignment with the ethics committee is key!

Is it possible to conduct an independent interventional study using a CE-marked medical device? ​

It depends. A study is possible if it is conducted within the device’s intended purpose, without additional invasive procedures, and after the CE-marked device has already been implemented at the health facility. In that case, no submission to the national competent authority is required and you already have access to the medical device, go for it!

However, the outlook is far less promising for all other independent research projects. If a study is planned outside the device’s intended purpose, with additional invasive procedures, or before the CE-marked MDSW is purchased, a study independent of the manufacturer is de facto not possible.

Assuming the product has already been implemented and can be controlled independently, or the manufacturer sells and implements the product but does not wish to provide further support for the study, the investigator faces the challenge of submitting the documents required by the national competent authority for the relevant “clinical investigation.” These include the clinical investigation plan (CIP) and the investigator’s brochure (IB). These documents require the following manufacturer-dependent content:

CIP (MDCG 2024-3):

  • Agreement between the sponsor and the manufacturer
  • List of the technical and functional features of the device
  • Risk management documents
  • Summary of the post market surveillance data

IB (MDCG 2024-5)

  • Internal technical documentation (software development, validation and testing documents, risk management, product classification, clinical evaluation, revision history)
  • Periodic safety update report or post market surveillance report

This effectively rules out conducting an independent interventional study here without the manufacturer’s involvement. If you want to know the details, you can check your national competent authority’s website to see when a clinical investigation needs to be submitted to them and whether any exemptions apply. For Germany, this information is available here (please note that the "Leitfaden zur regulatorischen Ein­ordnung klinischer Studien mit Medizin­produkten" does not cover in-house medical devices).